Interrogating the Logs
Each service writes its own log file:
| Service | Log file |
|---|---|
| DICOM2ORM | %ProgramData%\Flux Inc\DICOM7\DICOM2ORM\logs\dicom7-dicom2orm.log |
| DICOM2ORU | %ProgramData%\Flux Inc\DICOM7\DICOM2ORU\logs\dicom7-dicom2oru.log |
| ORM2DICOM | %ProgramData%\Flux Inc\DICOM7\ORM2DICOM\logs\dicom7-orm2dicom.log |
| ORU2DICOM | %ProgramData%\Flux Inc\DICOM7\ORU2DICOM\logs\dicom7-oru2dicom.log |
| ORU2REPORT | %ProgramData%\Flux Inc\DICOM7\ORU2REPORT\logs\dicom7-oru2report.log |
| HL72HL7 | %ProgramData%\Flux Inc\DICOM7\HL72HL7\logs\dicom7-hl72hl7.log |
If a service runs with --path, its log is in the logs folder below that path.
Rotation
- Each file grows to at most 10 MB, then a new file is started (
dicom7-<service>_001.log,_002.log, …). - The 7 most recent files are kept; older files are deleted automatically.
- The services log at Information level and above.
When the service runs from a console, the same entries are also written to the console.
Format
Each entry has a timestamp with UTC offset, a three-letter level, and the message:
text
2026-10-30 09:15:02.481 -04:00 [INF] Loading configuration from path: C:\ProgramData\Flux Inc\DICOM7\ORM2DICOM\config.yaml
2026-10-30 09:15:02.655 -04:00 [INF] Set cache folder to: C:\ProgramData\Flux Inc\DICOM7\ORM2DICOM\cacheLevels are INF (information), WRN (warning), ERR (error), and FTL (fatal: the service stopped).
Searching the logs
PowerShell:
powershell
cd "C:\ProgramData\Flux Inc\DICOM7\ORU2DICOM\logs"
# Warnings, errors and fatal entries
Select-String -Path dicom7-*.log -Pattern '\[(WRN|ERR|FTL)\]'
# Everything about one message control ID or accession number
Select-String -Path dicom7-*.log -Pattern 'ACC12345'
# Follow the live log
Get-Content dicom7-oru2dicom.log -Wait -Tail 50Command prompt:
cmd
findstr /C:"[ERR]" /C:"[FTL]" "C:\ProgramData\Flux Inc\DICOM7\DICOM2ORM\logs\dicom7-*.log"Useful entries
| Entry | Meaning |
|---|---|
Loading configuration from path: … | The configuration file the service used. |
Configuration file not found: … | No config.yaml at the expected path; the service is running on built-in defaults. |
Error loading configuration file | config.yaml could not be parsed; the service is running on built-in defaults. |
Set cache folder to: … | The cache folder in use. |
Failed to connect to HL7 receiver <host>:<port> | DICOM2ORM or DICOM2ORU could not reach the HL7 receiver. The message stays in the cache and is retried. |
HL7 server rejected message: … | The HL7 receiver answered with a response the sender treats as a rejection. |
No ACK received (timeout), but message was sent. Considering successful | The HL7 receiver did not answer in time; the sender treated the message as delivered. See the HL7 Conformance Statement. |
Empty response from receiver, but message was sent. Considering successful | The HL7 receiver closed without answering; the sender treated the message as delivered. |
… Please check the template and restart … | An HL7 template could not be loaded. Fix the template and restart the service. |
Capture-only mode is active … | HL72HL7 is storing messages but not forwarding them. |
Quarantined <n> abandoned staging records during startup | HL72HL7 found incomplete captures from an interrupted shutdown. See HL72HL7. |
Protected health information
The logs and the cache contain patient data from the messages and images that pass through the services. Restrict access to %ProgramData%\Flux Inc\DICOM7 and handle exported logs as protected information.