Skip to content

Installing DICOM7 ​

Prerequisites ​

  • 64-bit Windows
  • .NET Framework 4.6.2 or later
  • Administrator rights
  • Network access between the host and the DICOM and HL7 systems you are connecting, on the ports you configure
  • For ORU2REPORT only: Microsoft Edge or Google Chrome

Installers ​

DICOM7 is distributed as two Windows installers:

InstallerContents
DICOM7Setup-x64-<version>.exeThe DICOM7 suite. Lets you choose which services to install.
DICOM7-HL72HL7-Setup-x64-<version>.exeA standalone installer for the HL72HL7 broker only.

Contact Flux support to obtain the installer for your license.

Installing the suite ​

  1. Run DICOM7Setup-x64-<version>.exe as an administrator.

  2. Read the information page, then choose the installation folder (default C:\Program Files\Flux Inc\DICOM7).

  3. Choose the components to install:

    ComponentInstalled by FullWindows service
    Core Files (required)Yes—
    DICOM2ORM — DICOM Worklist to HL7 ORMYesDICOM7_DICOM2ORM
    DICOM2ORU — DICOM C-STORE to HL7 ORUYesDICOM7_DICOM2ORU
    ORM2DICOM — HL7 ORM to DICOM Modality WorklistYesDICOM7_ORM2DICOM
    ORU2DICOM — HL7 ORU to DICOM C-STOREYesDICOM7_ORU2DICOM
    ORU2REPORT — HL7 ORU to PDF/HTML reportsYesDICOM7_ORU2REPORT
    HL72HL7 — HL7-to-HL7 brokerNo (choose Custom)DICOM7_HL72HL7
  4. Finish the wizard. For each selected component the installer:

    • copies the program files to the installation folder,
    • creates %ProgramData%\Flux Inc\DICOM7\<SERVICE>\config.yaml from the default configuration only if it does not already exist,
    • registers the Windows service and starts it.

Install only what you need

The default HL7 listener port of ORM2DICOM, ORU2DICOM, and HL72HL7 is 7777. If you install more than one of them on the same host, give each its own HL7.ListenPort before starting them, otherwise only the first one to start can listen. See Troubleshooting.

The services start with the default configuration. Its hosts, ports, and AE titles are examples, not the endpoints agreed for your site. Configure each service, then restart it. See Configuration and Starting and Stopping the Services.

Installing the HL72HL7 broker on its own ​

  1. Run DICOM7-HL72HL7-Setup-x64-<version>.exe as an administrator.
  2. Choose the installation folder (default C:\Program Files\Flux Inc\DICOM7 HL72HL7 Broker).
  3. The installer:
    • copies the program files,
    • creates %ProgramData%\Flux Inc\DICOM7\HL72HL7\config.yaml only if it does not already exist,
    • restricts the data folder to SYSTEM and the local Administrators group,
    • validates the configuration with HL72HL7.exe --validate-config and stops the installation if it is invalid,
    • registers the DICOM7_HL72HL7 service and starts it.

A new HL72HL7 installation starts in CaptureOnly mode: it accepts and stores the allowed message types but does not forward them until you configure destinations. See HL72HL7.

Upgrading from ORM2HL7

The HL72HL7 installer replaces the older ORM2HL7 broker. When it finds the DICOM7_ORM2HL7 service it stops and removes it, removes the old program files, and installs HL72HL7 in its own folder. Configuration and cached data under %ProgramData% are not removed.

Default directories ​

PurposePath
Program files (suite)C:\Program Files\Flux Inc\DICOM7\
Program files (standalone HL72HL7)C:\Program Files\Flux Inc\DICOM7 HL72HL7 Broker\
Configuration%ProgramData%\Flux Inc\DICOM7\<SERVICE>\config.yaml
Cache%ProgramData%\Flux Inc\DICOM7\<SERVICE>\cache\ (configurable)
Logs%ProgramData%\Flux Inc\DICOM7\<SERVICE>\logs\
HL7 and report templates%ProgramData%\Flux Inc\DICOM7\<SERVICE>\

%ProgramData% is normally C:\ProgramData.

Service account ​

The services are registered to run as the Local System account and to start automatically with Windows. If a Windows service stops unexpectedly, Windows restarts it after 10 seconds, then after 20 seconds on a second failure.

ORU2REPORT launches a browser to render reports. Some browser installations do not render reliably under Local System. For a new ORU2REPORT installation, consider running DICOM7_ORU2REPORT under a dedicated, non-expiring account that can start the browser, and grant that account Modify access to C:\ProgramData\Flux Inc\DICOM7\ORU2REPORT. Upgrades keep the account you configured.

Verifying the installation ​

  1. Open services.msc and confirm each installed DICOM7_* service is Running.
  2. Open the service log in %ProgramData%\Flux Inc\DICOM7\<SERVICE>\logs\ and confirm that it reports the configuration it loaded and the ports it is listening on. See Interrogating the Logs.